Scheduled Maintenance
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 89
Subject to change without notice
7.2 Description of maintenance work
7.2.1 Visual control
MCS100FT
● Status line on the operator panel: Status field must light green.
If not: See logbook for cause.
● Logbook: Check for special entries.
● Sample gas outlet: Continuity, deposits.
● System cabinet: Visual check of the inside for
–Dryness
–Corrosion
–Unusual odor
–Unusual noises
Peripherals
● Sample gas sampling: Condition.
● Test gas supply: Condition, availability (use-by date), pressures.
7.2.2 Renewing the fan pad
The MCS100FT does not have to be switched off.
1 Remove the fan grill.
2 Insert a new fan pad immediately.
3 Refit the fan grill.
The interferometer contains a laser.
Do not open any parts of the interferometer enclosure.
Spare parts Part No.
Filter pad for cabinet fan 5309684
Filter pad air outlet (for version “IP54”) 5309684