Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 55
Subject to change without notice
5.7.7 Adjustment
Menu: MCS100FT/Adjustment
5.7.7.1 Automatically
Menu: MCS100FT/Adjustment/Automatically
This menu serves to start check and adjustment processes which then run automatically.
Touching displayed components allows to select only those components that have to be
checked.
To display all components again, touch: “activate all used components”.
Automatic O2 adjustment
Menu: MCS100FT/Adjustment/Automatically/Automatic O2 adjustment
IMPORTANT : Incorrect adjustment when O
2
sensor is dry
No dry gas may have been applied to the O
2
sensor for a longer period of time
before it is adjusted.
For example, perform the adjustment of the O
2
sensor before other compo-
nents are adjusted with dry gas.
Before exiting the menu, touch: “activate all used components”, otherwise it is
possible that not all components used are checked in automatic checks.
Designation Remark
O2 Sensor Driftcheck Drift check of O
2
sensor.
Presetting
p. 58, §5.7.7.3.
O2 Sensor Adjustment Adjustment check of O
2
sensor.
Presetting
p. 58, §5.7.7.3.
Automatically
Adjustment manual IR Components
Parameter
/MCS100FT/Adjustment
SCU
MCS100FT
Measure
p. 55, §5.7.7.1
p. 57, §5.7.7.2
p. 58, §5.7.7.3
/MCS100FT/Adjustment/Automatically/ Aut. O2 adjustment
SCU
MCS100FT
Measure
O2 Sensor Driftcheck
O2 Sensor Adjustment