88 MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG
Scheduled Maintenance
Subject to change without notice
7.1 Maintenance intervals
7.1.1 Recommended spare parts for 2 years operation
Maintenance work Reference W
1
1
W = weekly Q = quarterly, H = half-yearly, Y = yearly
Q
1
H
1
Y
1
System cabinet
Check measured values for plausibility ---- x x x x
Visual check of system cabinet
p. 89, §7.2.1 x x x
Replace filter pad for fan
p. 89, §7.2.2 x x x
Replace filter pad for air outlet
2
2
Only for the “IP54” version
p. 89, §7.2.2 x x x
Check drying agent cartridge
p. 90, §7.2.3 x x x
Maintenance by trained personnel  xx
Recommended spare parts Number Part No.
Filter pad for cabinet fan 8
1
1
As required
5309684
Filter pad for air outlet 8
2
2
Only for the “IP54” version as required
On request
Drying agent cartridge, interferometer
3
3
Recommendation: In case moisture has penetrated the interferometer.
1 5320799

Table of Contents