Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 33
Subject to change without notice
5.4 Measuring screen
Typical Measuring screen:
To exit the menu: Touch .
● Adjustment of Measuring screen
p. 40, §5.6.6
5.4.1 Status lines
The SCU has 2 status lines:
● First status line: SCU status line (higher level control unit).
● Second status line: Status line of the analyzer currently selected (MCS100FT or FID-
100FT)
The status lines have status fields (depending on parameter settings) to display the respec-
tive device status.
Significance of status fields
Abbrev. Color Significance Cause
None Green Proper operation ---
MReq, M
Yellow
Maintenance request A device function will shortly be restricted.
C Functional check Device-internal functional check running.
U Uncertain Uncertain device state.
F Red Failure Failure.
701
0.00
0.00 0.00
0.00
0.00
0.00 0.00
0.00
0.00
0.00
0.00
0.00
0.00
701
17.3
126
Measuring
5
Switch to higher menu level.
The “Measure” field is displayed
instead of Date and Time:
Touch “Measure” to return to the
Measuring screen.
Measured value box
p. 41, §5.6.6.1
Highlighted light brown:
The corresponding analyzer is
shown in the second status line
Date (dd.mm.yy) and time of the
operator panel (
§5.4.1)
Current menu level
2 status lines
p. 33, §5.4.1
/System Control Unit/Measuring/Measuring Screen 1
Component
Unit
SCU
MCS100FT
25.05.10
14:01
Measure
Component
Unit
Component
Unit
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
MCS100FT operating state with state
code (only internal significance)
Status fields (
p. 33, §5.4.1)
701
0.00
0.00 0.00
0.00
0.00
0.00 0.00
701
17.3
126
Status line of the analyzer
(MCS100FT or FID-100FT) for
which the measured value box
(
p. 41, §5.6.6.1) is activated
(highlighted light brown).
/System Control Unit/Measuring/Measuring Screen 1
Component
Unit
SCU
MCS100FT
25.05.10
14:01
Component
Unit
Component
Unit
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
SCU status line
(Parameter settings “SCU Techni-
cal Information“ manual)
Status fields
Top: SCU status field
Bottom: Analyzer status field