34 MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG
Operation
Subject to change without notice
Actions that can be taken when a status field is yellow or red:
Touch the colored measured value box: The associated analyzer is displayed in the sec-
ond status line.
If no analyzer shows an error: The cause is in the SCU.
Touch several times until the menu selection (
p. 35, §5.5) appears and then select
the menu of the analyzer or SCU involved.
Select the menu Diagnosis (depends on the analyzer).
Status fields presence and logic depend on the SCU parameter settings (
“SCU Technical Information“) or Analyzer manual.
The SCU status line parameter setting is normally “Group alarm”.
This means that the error message of an analyzer not displayed is also shown
as a status message in the SCU status line.
701
0.00
0.00 0.00
0.00
0.00
0.00 0.00
701
17.3
126
F
MReq
CU
F
M
C
U
/System Control Unit/Measuring/Measuring Screen 1
Component
Unit
SCU
MCS100FT
25.05.10
14:01
Component
Unit
Component
Unit
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
Measured value box status:
- White: Measured value OK
- Yellow: Maintenance request/uncertain
- Red: Failure
Status fields
Top: SCU status field
Bottom: Analyzer status field