Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 53
Subject to change without notice
5.7.6.1 Device parameters
Menu: MCS100FT/Parameters/Device Parameters
Device parameters
Device parameter times
An input screen is shown after touching a field.
Device parameter IR Cube
An input screen is shown after touching a field.
Designation Remark
Measurement Cell Optical Path Length Displays of optical length of sample gas cell
Designation Remark
Pre-Measuring Duration When Measuring mode is switched to: The speci-
fied period is defined as operating state “Pre-mea-
suring”. MCS100FT behavior (analog outputs, sta-
tus signals, etc) depends on the parameter
settings.
Back-Purging Duration Back purging cycle duration
Back-Purging, Period Interval in which the “pulse length” should occur.
Back-Purging, Pulse Length How long back purging runs.
Designation
1
1
IRC: IR Cube
Remark
IRC Operator Freely selectable
IRC Sample Name Freely selectable
IRC Sample Form Freely selectable
Device Parameters
Device Parameter Times
Device Parameter IR Cube
/MCS100FT/Parameter/Device Parameters
SCU
MCS100FT
Measure
p. 53
p. 53
p. 53
/MCS100FT/Par.../Device Parameters/Device Parameters
SCU
MCS100FT
Measure
8,48 mMeasurement Cell Optical Path Length
/MCS100FT/Par.../Device Par.../Device Par... Times
SCU
MCS100FT
Measure
170 sPre-Measuring Duration
10 s
180 sBack-Purging Duration
Back-Purging, Period
20 s
Back-Purging, Pulse Length
/MCS100FT/Par.../Device Par.../Device Par... IR Cube
SCU
MCS100FT
Measure
SICKIRC Operator
MeasureIRC Sample Name
IRC Sample Form Normal