66 MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG
Operation
Subject to change without notice
5.7.8.3 Drift check (QAL3) with span gas
Menu: MCS100FT/Diagnosis/Driftcheck (QAL3) with span gas
This menu serves to display the results of the drift check “with span gas”.
5.7.8.4 Driftcheck (QAL3) without span gas
Menu: MCS100FT/Diagnosis/Driftcheck (QAL3) without span gas
This menu serves to display the results of the drift check “without span gas”.
The menu is equivalent to menu Drift check (QAL3) with span gas (see above).
5.7.8.5 Energy values
Menu: MCS100FT/Diagnosis/Energy values
This menu serves to display 3 energies of the wavelength ranges.
To refresh the Table, per form “Upload all Parameters from Device” (
p. 51,
§5.7.4).
Designation Remark
MV Measured value number of component. Preset.
Name Name of component. Preset.
Dimension Unit of concentration. Preset.
Nominal Nominal concentration of reference value.
actual Actual concentration of the reference value last measured (without span gas)
actual Zero Actual concentration of the zero value last measured (with span gas)
Menu: MCS100FT/Diagnosis/Driftcheck (QAL3) with span gas
SCU
MCS100FT
Measure
MV Name Dimension Nominal actual actual Zero
MV1 HCl mg/m3 1.0E02 1.0E02 0.0E00
MV2 HF mg/m3 1.0E02 1.0E02 0.0E00
MV3 NH3 mg/m3 1.0E02 1.0E02 0.0E00
etc.
Mark Copy