MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 73
Subject to change without notice
5.8.5.5 Device parameters
Menu: FID-100FT/Parameter/Device parameters
This menu serves to display and enter device parameters.
5.8.5.6 Gas timing
Menu: FID-100FT/Parameter/Gas timing
This menu serves to set the purge times for automatic adjustment
Submenu Designation Remark
Device parameters Serial No. Serial number display
Fitting location Entry box for sampling point (fitting location)
name
System time System time Date and time display
Designation Remark
Sample gas delay Purge time [s] with sample gas after adjustment completion
until the measured value is released again.
Zero gas delay Purge time [s] with zero gas after zero gas activated during
adjustment. The averaging time then starts.
Zero gas average time Averaging time [s] for zero gas during zero point adjustment.
Span gas delay Purge time [s] with span gas after span gas activated during
adjustment. The averaging time then starts.
Span gas average time Averaging time [s] for span gas during span gas adjustment.
Purge time Purge time [s] with zero gas after adjustment completion until
the measured value is released again.
Device parameters
System time
/FID-100FT/Parameter/Device parameters
SCU
FID-100FT
Measure
/FID-100FT/Parameter/Gas timing
SCU
FID-100FT
Measure
Sample gas delay 120 s
Zero gas delay 300 s
Zero gas average time
Span gas delay
Span gas average time
Purge time
30 s
240 s
30 s
120 s