76 MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG
Operation
Subject to change without notice
5.8.7.2 Adjustment results
Menu: FID-100FT/Diagnosis/Adjustment results
Displays the adjustment results.
Designation Remark
Current adjustment Drift since “last adjustment” (top line in each case)
Last adjustment Drift since factory calibration (bottom line in each case)
Date Date of respective adjustment [jjjjmmtt]
Time Time of respective adjustment [hhmmss]
Zero drift abs. Absolute drift since last adjustment. Limit value as default. A
message occurs when exceeded
Span drift abs. Absolute drift since last adjustment. Limit value as default. A
message occurs when exceeded
Zero drift rel. Relative drift since last adjustment. Limit value as default. A
message occurs when exceeded
Span gas drift rel. Relative drift since last adjustment. Limit value as default. A
message occurs when exceeded
Span gas value Span gas concentration set.
(only shown for sensitivity drift)
Span gas name Span gas name.
(only shown for sensitivity drift)
Measurement range Full scale value
Measuring unit Sample gas unit.
Zero adjustment results
Responsivity adjustment results
/FID-100FT/Diagnosis/Adjustment results
SCU
FID-100FT
Measure
/MCS100FT/Adjustment/Adjustment results/xxxx
SCU
MCS100FT
Measure
Current adjustment
yyyymmdd
yyyymmdd
hhmmss
Last adjustment
Date
Time
etc.