Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 65
Subject to change without notice
Data logging: writing data Error when writing data Switch the MCS100FT off and on again.
Data logging: open file
Logbook error Error when writing into the
logbook
Flashcard missing No flashcard Please contact SICK Customer Service.
Extended
“X/E” classification in logbook, no display of further information
Spectral resolution warning Spectral resolution of inter-
ferometer is incorrect
Please contact SICK Customer Service.
Medium temperature warning Sample gas temperature out-
side tolerance
Check cause as far as possible.
Medium pressure warning Sample gas pressure errone-
ous
Medium flow warning Sample gas flow erroneous Probe/sample gas line clogged? Instru-
ment air supply available?
Measurement value range
warning
Measurement range
exceeded
- Sample gas concentration too high?
- Check sample gas feed for correct state
(cleanness, pressure, temperature).
- Feed zero gas and check display.
- Feed span gas and check display.
Measurement range x Measurement range x The MCS100FT has switched to measure-
ment range x
Dig.Out x Activation of valves ---
Uncertain
“U” classification in logbook, no display of further information
Spectral evaluation Error in calculation of spectra Please contact SICK Customer Service.
Numerical
Medium temperature out of
range
Sample gas temperature out-
side tolerance
Check cause as far as possible.
Medium pressure out of
range
Sample gas pressure outside
tolerance
Medium flow out of range Sample gas flow outside tol-
erance
Probe/sample gas line clogged? Instru-
ment air supply available?
Measurement value out of
range
Measurement range
exceeded
- Sample gas concentration too high?
- Check sample gas feed for correct state
(cleanness, pressure, temperature).
- Feed zero gas and check display.
- Feed span gas and check display.
1
If fault persists: Contact SICK Customer Service.
Logbook entry Description Possible cause/ clearance
1