Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 57
Subject to change without notice
5.7.7.2 Adjustment manual IR Components
Menu: MCS100FT/Adjustment/Adjustment manual IR Components
To terminate adjustment: Touch “Cancel” or go to “Measure”.
Designation Remark
Zero Adjust all Reset all IR components to zero.
The valves are switched automatically.
Man. adjust activate A menu for adjustment start is displayed (refer to below this Table).
Cancel Cancel the current check.
The respective valve is closed.
Test Gas For internal use only
“Operating state” Display of current operating state of the MCS100FT.
“1234” State active since [s].
“1234” Remaining time [s].
Designation Remark
Zero Adjust all (Originates from previous menu - without significance here)
Man. adjust activate (Originates from previous menu - without significance here)
Cancel Cancel the current check
The respective valve is closed.
Measured component Name of measured component
“1234” Actual measured value
Valve Checkmark: The valve for this component is open
Adjust. Referencepoint Set the factor of the reference point.
Execute this command only when you are sure that you wish to change the
factor.
Menu: MCS100FT/Adjustment/Adjustment manual IR Comp.
SCU
MCS100FT
Measure
Zero Adjust all
Man. adjust activate
Cancel
325 s
Measuring
23 s
Test Gas
/MCS100FT/.../Adjustment automatically IR components
SCU
MCS100FT
Measure
Zero Adjust all
Man. adjust activate
Cancel
1
HCl
Valve
9.23
Adjust. Referencepoint
2
HF
Valve
0.96
Adjust. Referencepoint