56 MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG
Operation
Subject to change without notice
Adjustment automatically IR components
Menu: MCS100FT/Adjustment/Automatically/Adjustment automatically IR components
Designation Remark
Zero Adjust all Set zero point for all activated components.
Driftcheck (QAL3) with
span gas
Drift check with span gas feed.
The MCS100FT is then in operating state “Drift check”.
Adjustment with span
gas
Adjustment of selected IR component with span gas.
Driftcheck (QAL3) with-
out span gas
[1]
1
Optional
Drift check without span gas feed.
The MCS100FT is then in operating state “Drift check”.
Adjustment without span
gas
[1]
Adjustment of selected IR component without span gas.
Cancel Cancel the current check
activate all used compo-
nents
Activation of all components used.
Component Name of component.
Active Checkmark: Component is active.
Operating state Display of current operating state of the MCS100FT.
1234 s” State active since [s].
1234 s” Remaining time [s].
/MCS100FT/.../Adjustment automatically IR components
SCU
MCS100FT
Measure
Zero Adjust all
Driftcheck (QAL3) with span gas
Adjustment with span gas
Driftcheck (QAL3) without span gas
Adjustment without span gas
Cancel
1
HF
2
32 s
Measure
HCl
activate all used components
Active
Active
127 s