Clearing Malfunctions
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 95
Subject to change without notice
8.3 Indicators on the operator panel
8.3.1 Status field red - “Stand-By“
When a malfunction occurs, MCS100FT switches automatically to “Stand-by.
In this state:
MCS100FT status field is red.
A status signal is set.
The sample gas path (including sampling probe) is purged with instrument air.
For FID-100FT: The FID-100FT fuel gas supply is switched off.
A relevant error message is displayed on the operator panel and entered in the logbook
(
p. 62, §5.7.8.2).
Correct the malfunction or have it corrected by trained personnel.
If it was possible to correct the malfunction without switching the MCS100FT off (e.g.
clearing a jam when the “sample gas flow” malfunction occurred), the MCS100FT auto-
matically switches back to Measuring mode.
8.3.2 “Current menu level” is red
When the current menu level is highlighted red:
The connection from MCS100FT resp. FID-100FT to SCU is interrupted.
1 Touch several times until the menu selection (
p. 35, §5.5) appears and then select
the menu of the analyzer or SCU involved.
The SCU and the MCS100FT or the FID-100FT regain contact.
If the connection is not established:
Check connection between MCS100FT resp. FID-100FT and SCU.
Restart the SCU (
p. 35, §5.5.1).
8.3.3 Time / date displayed is incorrect
When the time on the operator panel is wrong after every switch-on:
The battery in the SCU is empty.
Have the battery replaced by SICK Customer Service.
Time and date
p. 33, §5.4.1
701
0.00 0.00
000
000
000 000
701
17.3
126
F
F
/System Control Unit/Measuring/Measuring Screen 1
Component
Unit
SCU
Analyzer
25.05.10
14:01
Component
Unit
Component
Unit
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
NN
a.u.
Current menu level

Table of Contents