Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 69
Subject to change without notice
5.7.9 Maintenance
Menu: MCS100FT/Maintenance
5.7.9.1 Operation Mode Switch
Menu: MCS100FT/Maintenance/Operation Mode Switch
This menu serves to switch the MCS100FT to the desired operating state.
5.7.9.2 Status Reset
Menu: MCS100FT/Maintenance/Status Reset
All pending messages are reset.
Designation Remark
Measuring Switch the MCS100FT to measuring operation.
Maintenance Switch Maintenance mode on.
- A status signal ( delivered system documentation) is set.
- The yellow status indicator on the cabinet door lights.
- An entry is made in the logbook.
Back-Purge Backflushing runs once. Duration and cycle as per parameter settings in
p. 53, §5.7.6.1. Then automatic return to “Measuring”.
Purge MCS100FT sampling probe, sample gas line and gas path are purged with
instrument air.
A status signal ( delivered system documentation) is set.
Device State Shows the operating state.
State Active Since State Active Since
of For programs running automatically: Remaining time.
Designation Remark
Status Reset Reset all pending messages.
Response Result of reset.
Operation Mode Switch
Status Reset
/MCS100FT/Maintenance
SCU
MCS100FT
Measure
p. 69, §5.7.9.1
p. 69, §5.7.9.2
/MCS100FT/Maintenance/Operation Mode Switch
SCU
MCS100FT
Measure
Measuring
Maintenance
Purge
Back-Purge
14739 s
Device State Measure
State Active Since
0 sof
/MCS100FT/Maintenance/Status Reset
SCU
MCS100FT
Measure
Status Reset
Response