100 MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG
Technical Documentation
Subject to change without notice
9.3 Technical data
The Technical Data depend on the equipment on the MCS100FT.
The measuring components and measuring ranges can have individual settings deliv-
ered system documentation.
Components
Smallest measuring
range
Largest measuring
range
CH
4
0 ... 50 mg/m
3
0 ... 150 mg/m
3
CO 0 ... 75 mg/m
3
0 ... 1500 mg/m
3
HCl 0 ... 15 mg/m
3
0 ... 150 mg/m
3
HF 0 ... 3 mg/m
3
0 ... 10 mg/m
3
NH
3
0 ... 10 mg/m
3
0 ... 50 mg/m
3
NO 0 ... 200 mg/m
3
0 ... 2000 mg/m
3
N
2
O 0 ... 50 mg/m
3
0 ... 500 mg/m
3
NO
2
0 ... 100 mg/m
3
0 ... 500 mg/m
3
SO
2
0 ... 75 mg/m
3
0 ... 1500 mg/m
3
CO
2
0 ... 25 % by volume 0 ... 25 % by volume
C
3
H
8
0 ... 50 mg/m
3
0 ... 50 mg/m
3
H
2
O 0 ... 40 % by volume 0 ... 40 % by volume
O
2
0 ... 21 % by volume 0 ... 21 % by volume
TOC 0 ... 15 mg/m
3
0 ... 500 mg/m
3
Measured value recording
Detection limit: < 2% of full scale reading
Zero drift: < 3% of full scale reading per maintenance interval
For FID-100FT: < 2% of full scale reading per week
Sensitivity drift: < 3% of full scale reading per maintenance interval
For FID-100FT: < 2% of full scale reading per week
Temperature influence: < 2% of full scale reading / 10 K
Setting time T
90
: < 200 s
For FID: < 45 s
Limit values: 2 limit values as toggle
Device features
Optical path length of cell: 8.48 m (334 in.)
Laser: 1.5 mW , 633 nm
- Laser class 1: Outside interferometer
(radiation into cell)
- Laser class 3R: Inside interferometer
Device dimensions: 2100 x 800 x 600 (mm) (HxWxD) including 100 mm base
82.7 x 31.5 x 23.6 (in.) (HxWxD) incl. 3.4 in. Base
Weight: Approx. 260 kg (580 lb)
Enclosure color: RAL 7035
Sample gas volume, cell: 1.3 l (80 cu.in)
Heating temperature:
- Sample gas probe
- Sample gas line
- Cell
Max. 200°C (390°F)
Max. 200°C (390°F)
Max. 200°C (390°F)
Sample gas:
- Flow
- Temperature, cell
- Temperature, sampling point
- Input pressure
Approx. 300 l/h
Max. 200°C (390°F)
Max. 1300°C (2370°F)
90 ... 110 kPa (0.9 ... 1.1 bar)