72 MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG
Operation
Subject to change without notice
5.8.5.2 Measurement range
This menu serves to set the parameters for the measurement range.
5.8.5.3 Span gas setting
Menu: FID-100FT/Parameter/Span gas setting
This menu serves to enter the span gas name and concentration. Span gas concentration:
Approx. 80% of the measurement range set.
5.8.5.4 Sample gas name
Menu: FID-100FT/Parameter/Sample gas name
This menu serves to enter the sample gas name (freely selectable).
Designation Remark
Measurement range Full scale value (full scale value value is always “0”)
Unit ppm or mg org. C/m
3
/FID-100FT/Parameter/Measurement range
SCU
FID-100FT
Measure
Measurement range 15
Unit mg org C
/FID-100FT/Parameter/Span gas setting
SCU
FID-100FT
Measure
Span gas value 80ppm
Span gas name Propan
/FID-100FT/Parameter/Sample gas name
SCU
FID-100FT
Measure
Sample gas name CnHm