Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 63
Subject to change without notice
MCS100FT logbook entries
Device name Name of triggering unit.
Entries
1
Number of times errors have occurred.
Significance and default:
p. 54, §5.7.6.4
Text Logbook message (see below).
Class F = Failure
M = Maintenance request
C = Maintenance / Check
U = Uncertain
X/E = Extended
Date Begin Format: yy-mm-dd
For “Uncompressed”: Occurrence of message.
For “Compressed”: Last occurrence of message.
Time Begin Format: hh:mm:ss
For “Uncompressed”: Occurrence of message.
For “Compressed”: Last occurrence of message.
Date End Format: yy-mm-dd
For “Uncompressed”: Clearing of message.
For “Compressed”: Last clearing of message.
Time End Format: hh:mm:ss
For “Uncompressed”: Clearing of message.
For “Compressed”: Last clearing of message.
1
Only for compressed data storage
FID-100FT logbook entries
p. 77, §5.8.7.3
Logbook entry Description Possible cause/ clearance
1
Failure
“F” classification in logbook, status field on operator panel (
p. 33, §5.4) lights red
Unknown command Internal error. Switch the MCS100FT off and on again.
OS error
Bad Config.
File I/O
Numerical
Unknown failure
Numerical
Spectr. resolution out of
range
Spectral resolution is incor-
rect.
Please contact SICK Customer Service.
EEPROM EEPROM error Switch the MCS100FT off and on again.
IRCube comm. Communication with interfer-
ometer interrupted
Fuse in system cabinet (
p. 94, §8.1)
switched on?
Interferometer switched on? (Power plug
inserted correctly? Mains switch at bot-
tom of interferometer switched on). All 3
LEDs at the bottom of the interferometer
must light.
Check firm seating of the Ethernet con-
nection between the electronics and inter-
ferometer. The Ethernet LEDs at the elec-
tronics unit must light/flash.
Designation Remark