124 Clinitek Advantus Operator’s Guide
Specifications
Environmental Specifications
Electrical Requirements
Specification Value
Ambient Operating
Temperature
18°C to 30°C (64°F to 86°F).
At temperatures under 22°C (72°F), urobilinogen
and leukocyte results may be decreased, and at
temperatures above 26°C (79°F), increased.
Optimum Operating
Temperature
22°C to 26°C (72°F to 79°F).
Relative Humidity 20% to 80%, non-condensing, actively controlled
Optimum Relative
Humidity
35% to 55%
Indoor Use Only
Altitude up to 2000 meters
Ventilation 1709 BTU
IEC 1010-1 Installation Category II
IEC 1010-1 Equipment Classification Class I
IEC 1010-1 Pollution Degree 2
Requirement Value
Electrical Rating 100 to 240 VAC ± 10%
Power Requirements 50 to 60 Hz
Maximum Power Input 72 VA
Fuse Rating 2 A, 250 V, 2 AG, SB(T)
Line Leakage Current <0.5 mA in normal condition
<3.5 mA in single fault condition

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes

Related product manuals