Clinitek Advantus Operator’s Guide 25
Operating the System
Managing Results
Results are transmitted to the printer and computer as soon as all reagent
areas on the strip are read. If a record is flagged for a confirmatory report and
Edit flagged results is On, that record is not transmitted until after the end-of-
run reports complete.
End-of-Run Reports
The analyzer may display up to 3 end-of-run reports when the run, or a STAT
test, is completed. These reports display if you have marked any analytes to
flag for confirmatory or microscopic tests, and if Mark positives is On.
To request these reports:
1. Specify 1 or more tests for the Confirmatory Reports A and B or
Microscopy Report.
2. In the Setup routine, select
On for Edit flagged results.
The Confirmatory and Microscopic Report screens display the
SEQ # and
ID of the record, and the abbreviation for each positive analyte marked for
flagging.
Up to 5 records may be displayed on 1 screen.
3. Use
Move Up and Move Down to view additional records.
If both the Confirmatory and Microscopic Reports contain records, the
Confirmatory Reports display first.
4. Edit these results before exiting the Confirmatory Report. Refer to Editing
Results in the Confirmatory Reports‚ page 25.
5. Select Print to print a report.
6. Select
Return to Ready/Run to exit the report screen.
If an error is reported for 1 or more analytes, a report displays after the
Confirmatory and Microscopic Reports. This report displays last.
7. Retest any specimens listed.
Editing Results in the Confirmatory Reports
Use this procedure to edit the results of confirmatory testing.
1. During the end of run review, access the Confirmatory Report screens.
2. Select a record from the Confirmatory Report A screen.
The flagged positive test results display.

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes