22 Clinitek Advantus Operator’s Guide
Operating the System
3. If needed, enter or scan the color and clarity.
4. Select
Enter or scan the Enter code.
5. Repeat steps 2 to 4 for each specimen.
Editing a Loadlist
Use this procedure to make changes to the loadlist when initial entry is
complete.
1. Use
Move Up and Move Down to select the record to edit.
2. Edit the ID number.
NOTE: You cannot change or delete the ID number during Run mode.
Make any changes while the analyzer is in the Ready mode.
3. Select Delete to delete an item from the loadlist.
You can delete only the ID number being displayed or all IDs in memory.
4. Edit the color and clarity.
5. Select
Enter to accept the new number, color, and clarity.
Entering a Loadlist from a Host or Laboratory/Hospital System
You can connect the Clinitek Advantus analyzer to a host computer or
laboratory system. Refer to Appendix G, Computer and Printer Interface, for
more information.
1. Before sending a loadlist from a host or laboratory system, ensure that the
following conditions are true:
The analyzer is at the Ready/Run screen.
No IDs from an earlier loadlist are still stored in the analyzer. If a loadlist
was sent but is no longer needed, you can overwrite the unused IDs
with a new loadlist.
The computer port is set to computer port, ethernet port, or both.
The computer port options for Baud, Data, and Parity are correct for the
computer or LIS/HIS sending the loadlist. Refer to the specifications
accompanying the computer or Laboratory/Hospital Information System
for information on the required parameters.
The output format for the computer port is CCS. Refer to Computer Port
Options‚ page 104 for more information on setting the computer port.
NOTE: Loadlist data is only transferred if it is formatted correctly. If a
loadlist is not transferred, refer to Section 6, Troubleshooting, for possible
causes.

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes

Related product manuals