Clinitek Advantus Operator’s Guide 27
Operating the System
6. Select Enter to enter the data.
7. Repeat this procedure to add all required microscopic test data.
8. Select
Print to print the microscopic result data and the results of the
patient test on the analyzer.
9. Select Merge to store the microscopic data with the analyzer results.
When the test results are recalled, Microscopics displays on the results
display to show that microscopic results are stored with the test results.
Recalling Results
Up to 500 patient records and 200 quality control records are stored in
memory. Use the following procedure to recall 1 or more records:
1. At the Ready/Run screen, select Menu.
2. Select
Memory.
3. Recall a group of records.
all patient records
all QC records
the last batch of patient results
stored results by Patient ID
The number of records in memory displays next to the first 2 options.
The last batch of patient results tests are those tests run between the last
pause in testing and the latest test. If the latest test is a QC test, it is not
recalled.
4. If you selected
Search for stored results, enter the patient ID.
5. Select
Enter to start the search.
The earliest record of the selected group displays. The date and time the
record was stored displays, along with the Technician ID, SEQ #, and ID for
the record. All results are then listed. Positive results are flagged with an
asterisk (*) and edited results with an exclamation point (!).
6. Locate the first record to review using the movement keys shown on the
display.
The next lower- or higher-numbered record in memory is recalled when
Move Up and Move Down are used. The record 10 higher or lower is
recalled when Move Up 10 and Move Down 10 are used.
If microscopic results are merged with the patient test results,
Microscopics displays on the patient record.
7. Select
Microscopics to view the merged microscopic results.

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes