98 Clinitek Advantus Operator’s Guide
System Configuration
Setup Menu 6
At the fifth Setup menu, select Next Screen to access the sixth Setup menu.
Positive Levels for COL/CLA
Use this procedure to set the lowest positive result for color and clarity.
The analyzer uses these levels to determine which specimens meet the criteria
for the confirmatory and microscopic reports.
If Mark positives is On, positive results are marked with an asterisk (*) in the
displayed and printed report and in the data transferred to a host computer.
1. Select
Positive levels for COL/CLA.
The lowest level considered positive for color and clarity displays. The
available choices are those set earlier.
2. For each setting, use the cycle key to set the first level marked as positive.
All results later in the list are also marked positive.
3. Select
Previous Screen.
Flags for Confirmatory Test A
The confirmatory reports list those specimens that may require confirmatory
testing. Mark positives must be On to obtain the reports.
Use this procedure to select up to 5 tests for confirmatory report A.
1. Select
Flags for confirmatory test A to display a list of reported tests.
2. Select the box next to the tests to include in the confirmatory report.
A check mark displays in the box.
Select the box again to remove the check mark.
3. Select
Previous Screen.
Menu Option Default
Positive levels for COL/CLA N/A
Flags for confirmatory test A N/A
Flags for confirmatory test B N/A
Flags for microscopics N/A
Set QC options N/A