Clinitek Advantus Operator’s Guide 69
File Management
7 File Management
The analyzer stores the operating parameters, including those selected by the
user, and up to 500 patient results and 200 quality control results. The analyzer
automatically overwrites the oldest results when it exceeds capacity.
If you connect the analyzer to a computer, it automatically transfers results at
the end of a run. Refer to Section 8, System Configuration for information on
connecting to a computer.
If your analyzer is connected to a computer, or if you accidentally set Computer
port to Computer port, Ethernet port, or Both, the computer expects an
acknowledgement after it attempts to send results. If it does not receive the
acknowledgement, it continues to hold results in memory instead of overwriting
them. The analyzer generates an error when the memory nears capacity.
Refer to Section 6, Troubleshooting for error message details, and suggested
corrective actions.

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes