20 Clinitek Advantus Operator’s Guide
Operating the System
6. Place the reagent strip onto the supports of the strip loading station, with
the reagent pads facing up.
Place the strip to the right of and parallel to the push bar. Ensure that the
end of the strip is against the back wall of the platform and that it is not
touching the bottom of the strip loading station.
Figure 6 Placement of Reagent Strip
7. Repeat steps 2 to 6 for each specimen.
When the push bar is to the far left of the platform, you can place a new strip
on the loading station until the previous strip placed enters the waste bin.
When the final strip moves to the waste bin, the run ends, and end of run
reports are processed.
Refer to Printing and Transmitting Results‚ page 28 for information on printing
and transmitting the results.
CAUTION
Improper placement may cause the analyzer to jam or the strip to
incorrectly align under the readheads.

Table of Contents

Other manuals for Siemens Clinitek Advantus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens Clinitek Advantus and is the answer not in the manual?

Siemens Clinitek Advantus Specifications

General IconGeneral
Test TypesUrinalysis
InterfaceRS-232, USB
CalibrationAutomatic calibration
Measurement PrincipleReflectance photometry
Sample TypeUrine
Power Supply100-240 VAC, 50/60 Hz
TypeUrine analyzer
DisplayTouchscreen
Reagent TypeTest strips
Measurement ParametersGlucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes

Related product manuals