Clinitek Advantus Operator’s Guide 73
System Configuration
6. Place the analyzer on a firm, level work surface in the designated work
area.
7. Ensure that the analyzer is level, and that the back and sides of the
analyzer are at least 7.6 cm (3 in) from any adjacent wall or analyzer.
8. Locate the piece of foam packing that is under the read area cover.
The foam has a red tag attached.
9. Remove the foam by gently pulling the red tag down and forward.
Figure 22 Remove the Foam