EasyManua.ls Logo

Siemens SITRANS T Series - How TR320 Works

Siemens SITRANS T Series
236 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.3.7 How TR320 works
The functional principle is described below based on the function chart.
4
1
3
2
3 (+)
4 (-)
µC1
µC2
4 ... 20 mA U, I
out
SITRANS TR320
D
A
A
D
1 (+)
2 (-)
LED Galvanic isolation
Test terminals Input
Output A/D Analog-to-digital converter
μC2 Microcontroller, on the pri‐
mary side
μC1 Microcontroller, on the secondary side
D/A Digital-to-analog converter
U Supply voltage
I
out
Output current
Figure 4-14 Function chart for SITRANS TR320
Input for connecting di󹪜erent signal sources. Section: Input (Page 201).
This signal is converted to a digital signal in an analog-to-digital converter (A/D).
The digital signal is evaluated in a microcontroller (μC1) on the secondary side and corrected
to match the input characteristic.
The digital signal is transferred via the galvanic isolation to the microcontroller (μC2) on
the primary side.
The analog output value is calculated in the microcontroller (μC2) on the primary side. The
functional status is indicated by LED and the communications data is prepared.
The digital-to-analog converter (D/A) subsequently converts the signal into the output
current of 4 to 20 mA.
Description
4.3 SITRANS TR320/TR420
SITRANS TH100/TH320/TH420/TR320/TR420/TF320/TF420 with 4 to 20 mA/HART
48 Operating Instructions, 05/2020, A5E41864807-AB

Table of Contents

Other manuals for Siemens SITRANS T Series

Related product manuals