4.5 USB modem and SIPROM T
4.5.1 Applications
Use the USB modem only for the purposes speci󹪝ed in these instructions.
The USB modem with SIPROM T parameter assignment software is used for parameter
assignment and operation of the following temperature transmitters:
SITRANS TH100Slim/TH100/TH200/TH320 with 4 to 20 mA
SITRANS TR200/TR320 with 4 to 20 mA
SITRANS TF with SITRANS TH200
SITRANS TF320 with 4 to 20 mA
Connect the temperature transmitter to the PC via the USB modem. The required supply voltage
of the temperature transmitter is provided via the USB modem.
NOTICE
Improper use of the USB modem
The USB modem and the connected devices can be damaged.
Only use the USB modem for parameter assignment of the named Siemens temperature
transmitters.
Always use the SIPROM T parameter assignment software.
Refer to the information in section Technical data (Page 193).
Observe the technical data of the temperature transmitters in the associated operating
instructions. You can 󹪝nd the operating instructions on the Internet at Instructions and
manuals (http://www.siemens.com/processinstrumentation/documentation).
4.5.2 Product features
USB port (USB V1.1, compatible with USB 2.0)
Galvanic isolation between the PC and the temperature transmitter with parameters to be
assigned
Adherence to the Ex requirements for the connected temperature transmitters
Feeding of the USB modem with supply voltage directly from the USB port of the PC
Description
4.5 USB modem and SIPROM T
SITRANS TH100/TH320/TH420/TR320/TR420/TF320/TF420 with 4 to 20 mA/HART
58 Operating Instructions, 05/2020, A5E41864807-AB