Performance Verification
F-12
VX4792 User Manual
a. Connect the oscilloscope: Connect the WAVEFORM output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope.
b. Set the oscilloscope controls:
Vertical CH1
CH1 coupling DC
CH1 scale 0.2 V/div
CH1 input impedance: 50 W
Horizontal
Sweep 5 ms/div
Trigger
Source CH1
Coupling DC
Slope Positive
Level –100 mV
Mode Auto
2. To set the VX4792 Arbitrary Waveform Generator controls and select the
waveform file, type:
  
  
  
3. Check against limits: Check that the oscilloscope displays a five-cycle sine
wave that has an amplitude of five divisions.
4. End procedure: Disconnect the oscilloscope.
This procedure checks the operation of the Triggered mode.
Electrical Characteristic Checked: Operating mode, Triggered, on page E–4.
Equipment Required: Two 50 W coaxial cables, a function generator, and an
oscilloscope.
Prerequisites: The instrument must meet the prerequisites listed on page F–3.
Procedure:
1. Install the test setup and set test equipment controls (see Figure F–4):
Check Triggered Mode