Performance Verification
VX4792 User Manual
4. Setup controls of additional VX4792 Arbitrary Waveform Generator modules
(if installed):
a. To set the third VX4792 Arbitrary Waveform Generator controls, type:
b. To set the fourth VX4792 Arbitrary Waveform Generator controls, type:
c. To set the fifth VX4792 Arbitrary Waveform Generator controls, type:
d. To set the sixth VX4792 Arbitrary Waveform Generator controls, type:
5. To begin the test, type:
6. Check skew between the first, second, third and fourth modules: Check that
the skew (time variance) between the rising edges of the waveforms
displayed on the oscilloscope is within 4 ns (see Figure F–20).