7-12
764333-766
Figure 7-28. Disconnect Hydraulic Lines From
Main Valve Block
12. Disconnect electrical connections.
13. Remove two fastening screws and replace main
valve block.
14. Reassemble by reversing procedure.
NOTE: When reassembling the hydraulic main valve
block, note the following:
1) Ensure all screws are secured with Loctite
®1
242.
2) Replace (do not reuse) hydraulic screws and sealing
washers (see Figure 7-28). Ensure bolt threads,
manifold threads and hose connectors are free of
rubber seal particles.
3) Torque new screws to 14 ± 1.5 N•m (124 ±
13 lbf•in).
4) Follow 4085 Surgical Table Hydraulics figure in
the IPB Section of this manual to ensure proper
reconnection.
5) After reassembly, ensure all table functions operate
properly. Exercise all articulations at least five
times to completely purge system of trapped air.
Adjust movement speed if necessary (refer to
S
ECTION 6.6, TABLE ARTICULATION TIMES).
6) Check hydraulic oil level with table completely
lowered and tabletop level. Level should be 2-9/16"
(65 mm). Add oil (P764322-636) as needed.
1
Loctite is a registered trademark of Loctite Corporation, a
Henkel Company.
7.20 REMOVE TABLETOP SECTIONS
Remove the Tabletop Sections as follows:
NOTE: Minor adjustments to accessory receptacle positions
may be made by loosening tabletop screws, inserting the
accessory and then torque the screws to 6 N•m (53.1 lbf•in).
7.20.1 Seat Section
Remove Tabletop Seat Section as follows:
1. Remove and save six slotted pan head screws
fastening tabletop to frame (see Figure 7-29).
2. Remove and set aside in safe area to avoid damage.
3. Reassemble in reverse order.
Figure 7-29. Remove Tabletop Seat Section
7.20.2 Kidney Section
Remove Tabletop Kidney Section as follows:
1. Remove and save four slotted pan head screws
fastening tabletop to frame (see Figure 7-30).
2. Remove and set aside in safe area to avoid damage.
3. Reassemble in reverse order.
Figure 7-30. Remove Tabletop Kidney Section
7.20.3 Back Section
Remove Tabletop Back Section as follows:
1. Remove and save 12 slotted pan head screws
fastening tabletop to frame (see Figure 7-31).
2. Remove and set aside in safe area to avoid damage.
3. Reassemble in reverse order.
Figure 7-31. Remove Tabletop Back Section
Lower
Fastening
Screw
Two-Way Hydraulic
Screw and Sealing
Washers
Hydraulic
Screw
and
Sealing
Washers
Upper
Fastening
Screw
Screws
Screw
Screws
Screws

Table of Contents

Other manuals for Steris 4085

Questions and Answers:

Steris 4085 Specifications

General IconGeneral
Table TypeGeneral Surgical Table
Trendelenburg30°
Reverse Trendelenburg30°
Lateral Tilt20°
Table Top Dimensions20 in (51 cm) wide
Kidney Elevator4 in (10 cm) lift
Weight Capacity1, 000 lbs (454 kg)