6-3
764333-766
22. Display shows BACK DOWN MAX. Lower back
section to maximum position and press Reverse
Orientation button to save setting.
23. Display shows BACK UP MAX. Raise back
section to maximum position and press Reverse
Orientation button to save setting.
24. Display shows SLIDE HEAD MAX. Use BACK
DOWN button to return table to center. Do not
use LEVEL button.
25. Slide tabletop to maximum head position then
press Reverse Orientation button to save setting.
26. Display shows SLIDE FOOT MAX. Slide tabletop
to maximum foot position then press Reverse
Orientation button to save setting. Slide should
hesitate momentarily at zero position.
27. Display shows TABLE DOWN MAX. Lower
tabletop to lowest position then press Reverse
Orientation button to save setting.
28. Display shows TABLE UP MAX. Raise tabletop
to highest position then press Reverse
Orientation button to save setting.
29. Display shows SAVE VALUES. Press Reverse
Orientation button to reset all sensor alarms and
save calibration values.
NOTE: Calibration data is lost if you fail to press
Reverse Orientation button in this step.
Important: Do not press the LEVEL button until
Step 31 is performed.
30. Hand Control exits ADJUST MODE.
31. Press STOP button on Primary Hand Control to
turn power OFF then press any button (but STOP)
to return power to table.
32. Verify table returns to level correctly from all
positions, including Flex and Reflex.
33. Verify all ALS functions.
6.3.2 Newer Control Board (P150832-769)
1. Plug table into appropriate facility outlet.
2. Press any primary hand control button to turn table
ON.
3. Engage floor locks.
4. Ensure leg section is attached to table.
5. Raise table high enough to get a full leg down
articulation without hitting table base.
6. Simultaneously press and release LEVEL and
Reverse Orientation buttons.
7. Display shows ADJUST MODE.
NOTE: During table sensor calibration, note the
following:
1) The Reverse Patient LED must be OFF.
2) Do not press LEVEL button during procedure
or control exits Calibration Mode without
storing the values.
3) Pressing FLOOR LOCK button during
procedure orders the control to skip displayed
articulation. This is NOT recommended during
a final calibration.
8. If ADJUST MODE does not display or Reverse
Patient LED is ON, turn Hand Control OFF (press
STOP) then ON and repeat Step 6 until correct
results appear.
9. Press Reverse Orientation button, display shows
LEVEL ADJ.
10. Using a level, position tabletop (starting with seat
section) so all sections are level in both horizontal
directions.
11. Slide tabletop until it is approximately 3/16"
(4.8 mm) past foot end of tilt frame (see
Figure 6-2). This is known as "Zero Slide" position.
12. PressReverse Orientation button to save setting.
13. With display showing SLIDE FOOT MAX, slide
tabletop to maximum foot end position.
14. Press Reverse Orientation button to save setting.
15. Display shows SLIDE HEAD MAX. Slide
tabletop maximum head end position.
16. Press Reverse Orientation button to save setting.
Display shows TABLE DOWN MAX.
17. Use SLIDE FOOT button to return table to center.
Do not use LEVEL button.
18. With display showing TABLE DOWN MAX,
lower table to maximum down position.
19. Press Reverse Orientation button to save setting.
20. Display shows TABLE UP MAX. Raise table to
maximum up position.
21. Press Reverse Orientation button to save setting.
Display shows REV TREND MAX.
22. Use HEIGHT DN button to return table to near
middle height. Do not use LEVEL button.
23. With display showing REV TREND MAX
, move
tabletop to maximum reverse Trendelenburg
position.
24. Press Reverse Orientation button to save setting.
25. Display shows TREND MAX. move tabletop to
maximum Trendelenburg position.