6-2
150832-639 Operator Manual Troubleshooting
* Service charges may be incurred. Consult your warranty for details.
** Use optional foot control or auxiliary (override) control system if hand control is faulty [refer to SECTION 4.5, BACKUP
(OVERRIDE) CONTROL SYSTEM].
Table 6-1. Troubleshooting Chart
SITUATION ACTION
Hand Control does not activate when table is plugged
into electrical outlet and any pushbutton (except
STOP) is pressed .
Main battery fuse blown – replace.
Faulty power cord – replace.
Hand control not plugged in properly – correct.
Faulty hand control** – replace.
Call STERIS.*
Table remains on battery power even when main
power cord is plugged into facility power.
Incorrect power cord connection – correct.
Faulty power cord – replace.
Main fuse blown – replace.
Call STERIS.*
No functions using hand control. Hand Control not plugged in properly – correct.
No Power – table deactivates after 30 minutes of non-
use. Press any button on hand control (except
STOP).
Faulty hand control** – replace.
Call STERIS.*
Table does not tilt right/left using hand control. Hand control not on – Press any button on hand
control (except STOP).
Faulty Hand control** – replace.
Call STERIS.*
Table does not Trendeleburg/Reverse Trendeleburg
using hand control.
Hand control not on – Press any button on hand
control (except STOP).
Faulty Hand Control** – replace.
Call STERIS.*
Table does not raise/lower using hand control. Hand control not on – Press any button on hand
control (except STOP).
Faulty hand control** – replace.
Call STERIS.*