5-4
150832-639 Operator Manual Routine Maintenance
5. Turn hand control OFF and clean as follows:
NOTE: Do NOT immerse hand control in liquid.
a. Hold can 6-8" (152-203 mm) from surface and spray
cleaning fluid liberally on hand control.
b. Wipe sprayed surfaces with a clean, lint-free cloth
dampened with water to remove cleaning fluid (dampening
cloth minimizes streaking).
5.2.3 End-of-Day Cleaning
Procedure
At the end of each day, perform the cleaning procedures as outlined
in S
ECTION 5.2.2, AFTER EACH USAGE.
5.2.4 Weekly Cleaning
Procedure
After each weekly use of the STERIS 4085 General Surgical Table,
clean/disinfect table as follows:
1. Perform Steps 1 through 4 of S
ECTION 5.2.2, AFTER EACH
U
SAGE.
2. Check table casters and floor locks for any accumulated debris
and clean as follows:
a. Ensure floor locks are properly engaged (raising casters off
floor).
b. Hold cleaner spray can 6-8" (152-203 mm) from caster and
spray cleaning fluid liberally on the caster.
c. Wipe caster with a cloth, dampened with water, to remove
cleaning fluid and debris.
d. Perform Steps b and c for remaining three casters.
3. Articulate table through all movements and clean all additional
surfaces exposed during these articulations as follows:
a. Hold spray cleaner can 6-8" (152-203 mm) from surface to
be cleaned and spray cleaning fluid liberally on surface.
b. Wipe sprayed surfaces with a clean, lint-free cloth,
dampened with water to remove cleaning fluid.
c. Wipe cleaned surfaces again with a clean, damp, lint-free
cloth to remove any remaining residue.
4. Turn hand control OFF when finished with cleaning procedure.
5. Perform Steps 5 and 6 of S
ECTION 5.2.2, AFTER EACH USAGE.
5.3 Bi-Weekly
Maintenance
Operate each STERIS 4085 General Surgical Table function (refer to
S
ECTION 4, OPERATING INSTRUCTIONS). Operation must be smooth
and quiet. If table operation is not smooth, have a qualified
technician repair the table. Never permit inexperienced, unqualified
persons to attempt to make any repairs to the table.
5.4 Monthly
Maintenance
Check STERIS 4085 General Surgical Table casters and floor locks
for any accumulated debris and clean as follows:
NOTE: Ensure Hand Control is in proper working order and good
condition.
1. Ensure floor locks are properly engaged.
2. Hold spray cleaner can 6-8" (152-203 mm) from caster and
spray cleaning fluid liberally on the caster.