5-10
150832-639 Operator Manual Routine Maintenance
5.9 LEVEL Check Check the STERIS 4085 General Surgical Table tabletop LEVEL
function as follows:
1. Review and follow instructions presented in S
ECTION 4,
O
PERATING TABLE and prepare STERIS 4085 General Surgical
Table for operation.
2. Press LEVEL pushbutton.
NOTE: When operating in REVERSE ORIENTATION, LEVEL
function does NOT include the back section for compliance with
anesthesia procedures. Ensure REV. ORIENT is not active.
3. When depressed, tabletop is gradually returned to level or
horizontal position by aligning tabletop sections in following
order:
a. KIDNEY Elevator and lateral tilt.
b. Trendelenburg and back.
c. Leg.
4. Use a level and ensure tabletop is level. If not, call STERIS.
5.10 Auxiliary Control
Check
Check the STERIS 4085 General Surgical Table auxiliary control
system as follows:
1. Review and follow instructions presented in S
ECTION 4, BACKUP
(O
VERRIDE) CONTROL SYSTEM and prepare the STERIS 4085
General Surgical Table for auxiliary control system operation.
2. With tabletop at extreme low height, press HEIGHT UP
pushbutton and pump pedal.
3. Ensure tabletop raises correctly. If not, call STERIS.