150832-639 Operating Manual Safety Precautions
WARNING – DISPOSAL HAZARD:
CAUTION POSSIBLE EQUIPMENT DAM AGE:
REMEMBER – POSSIBLE TIME DELAY:
This product contains materials which may require disposal through appropriately licensed and
permitted hazardous waste management firms.
When moving the table to or from point of use, roll it carefully at moderate speed and only over smooth
floors. Maximum floor clearance is 1/4" (6.4 mm). Avoid door and elevator jambs, and obstructions
greater than 1/4" (6.4 mm). If necessary, lift table over obstructions, onto trucks, etc. Lift table evenly
and only by the table base. DO NOT transport articles (including accessories) on top of the table, and
DO NOT use a forklift to move the uncrated table.
Hang the hand control from side rail of the table when not in use to avoid possible damage to the
control.
When cleaning/disinfecting table, thoroughly read the cleaning fluid directions for use and follow all
directions and cautions as shown.
Do not spray cleaning fluid into electric receptacles and avoid spraying directly on emergency backup
buttons or into clearance space. Spray or drippage may settle onto electric circuits inside table causing
corrosion and loss of function.
Cleaning procedures requiring articulation of the table should be performed only by persons familiar
with table operation.
Route the hand control cord clear of any pinch points where cord could be damaged.
Use of incorrect hydraulic oil may severely damage the table and/or cause malfunction. Contact STERIS
for proper hydraulic oil.
After performing cleaning procedures, ensure pads, tabletop and X-ray tops are completely dry before
reinstalling. Moisture trapped between pads and X-ray tops may contribute to equipment damage, such
as X-ray top warpage.
Appropriate components of this surgical table have been tested and found in compliance with IEC
60601-1-2:2001, Medical Electrical Equipment – Part 1; General Requirements for Safety;
Electromagnetic Compatibility (EMC). There is, however, a potential for electromagnetic or other
interference between this equipment and other devices. Should user experience interference, relocate
this device or minimize the use of the affected equipment while this device is in operation.
Insufficient clearance space will make repairs more difficult and time-consuming. Refer to the
equipment drawing for minimum clearance for service/maintenance access requirements.