4-5
Operating Instructions Operator Manual 150832-639
BACK UP: When depressed, the tabletop back section raises.
BACK DN: When depressed, the tabletop back section lowers.
LEG UP: When depressed, the tabletop leg section raises.
LEG DN: When depressed, the tabletop leg section lowers.
KIDNEY UP: When depressed, the tabletop Kidney™ Elevator
Mechanism raises to the maximum 4" (110 mm). The LCD Display
Kidney Bridge (refer to Table 1-3) ICON displays.
KIDNEY DN:
When depressed, the tabletop Kidney Elevator
Mechanism lowers. The LCD Display Kidney Bridge (refer to
Table 1-3
)
ICON is no longer displayed.
LEVEL: When depressed, the tabletop is gradually returned to level
or horizontal position by aligning the tabletop sections in the
following order:
Kidney Elevator Mechanism and lateral tilt;
Trendelenburg and back;
•Leg.
NOTE: When operating in Reverse Orientation, LEVEL function does
NOT operate.
FLOOR LOCK: When FLOOR LOCK button is depressed, the LCD
Display shows the ICONs for Locked (green) or Unlocked (orange).
To lock table, press and hold FLOOR LOCK button. LCD
Display shows “Hold 1 sec to Lock.” After one second, control
system orders floor locks to lower. While floor locks are lowering,
LCD Display shows “LOCKING.” Once table is locked, Locked
(green) ICON appears on LCD Display and green LED lights at
FLOOR LOCK pushbutton.
To unlock table, press and hold FLOOR LOCK button. LCD
Display shows “Hold 3 secs to Unlock.” After three seconds,
control system orders floor locks to raise. While floor locks are
raising, LCD Display shows “UNLOCKING.” Once table is
unlocked, UnLocked (orange) ICON appears on LCD Display
and amber LED lights at FLOOR LOCK pushbutton.
NOTE: When table is in the UNLOCK configuration, NO tabletop
functioning is active. If a tabletop function button is pressed while
table is in the UNLOCK configuration, the LCD Display shows LOCK
TABLE.
STOP: When depressed, all hand control function buttons
deactivate. All functions are stopped immediately.
NOTE: The hand control automatically deactivates upon nonuse after
30 minutes if floor locks are released and 10 hours when floor locks
are engaged. The hand control is reactivated by depressing ANY
button but STOP.