4-3
764333-766
Container of clean water.
Cleaning products listed in Table 4-3,
R
ECOMMENDED CLEANING PRODUCTS.
NOTE: When using disinfectants, follow all
manufacturer's label recommendations.
Check all cables for damage or fraying. 1x per year
Inspect electrical cord and Hand Control cable. 1x per year
Verify operation of Backup (Override) Hand Control. 1x per year
Verify condition of batteries and battery charger status. 1x per year
Verify Auto Limit Sensor status. 1x per year
Measure voltage of both batteries. Difference in readings must be less than 0.2 V. Perform
articulations for at least two minutes. Voltage must not drop more than 0.6 V. Wait two
minutes and verify voltage has risen by 0.3 V at most. Supply 120 V and ensure voltage
rises above 28 V.
1x per year
6.0 TABLETOP OPERATION
Check operation of LEVEL function (also check Hand Control LCD Display function). 1x per year
Check operation of back section (also check Hand Control LCD Display function). 1x per year
Check operation of leg section (also check Hand Control LCD Display function). See
S
ECTION 4.10, LEG SECTION CHECKS.
1x per year
Check TILT LEFT and TILT RIGHT operation (also check Hand Control LCD Display
function).
1x per year
Check TREND, then REV. TREND operation (also check Hand Control LCD Display
function).
1x per year
Check HEIGHT UP and HEIGHT DOWN operation (also check Hand Control LCD
Display function).
1x per year
Check headrest section operation. 1x per year
Check SLIDE function (also check LCD Display). 1x per year
7.0 TABLE RIGIDITY
Verify head and leg sections locking devices. 1x per year
Check for any play in floor-lock legs. 1x per year
8.0 FINAL TEST
Examine all lubricated parts. Apply lubrication (Lubriplate No. 2, R6400-826) as needed. 1x per year
Reinstall any pads that were removed. Check for rips, tears, etc. 1 x per year
Secure and properly seal all covers and shrouds. 1 x per year
Inspect area to ensure removal of all materials used during inspection. 1 x per year
Table 4-1. Preventive Maintenance Schedule for the STERIS 4085 General Surgical Table