4-9
Washer Operation Operator Manual 122997-341
>> If sump flush option is available, refreshed water still
present in sump is gravity drained. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 128.5F
FLUSHDRAIN T= 2:00
ACID WASH PHASE
>> Hot water, from building supply (or from acid wash tank, if
option applies) fills sump. If temperature is guaranteed,
actual sump temperature alternates with guaranteed set
temperature. Display shows:
CYCLE 10 (G=) 137.4F
ACIDWASHFILL T= 2:00
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time only counts
down when solution temperature is equal or greater than set-
point.
>> Water is heated while recirculating (CONDUCTIVITY
OPTION ONLY). If temperature is guaranteed, actual sump
temperature alternates with guaranteed set temperature.
Display shows:
CYCLE 10 (G=) 141.5F
RECIRC/HEATING T= 5:00
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time only counts
down when solution temperature is equal or greater than set-
point.
>> Acid detergent is injected into sump. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 141.5F
INJECTINGACID T= 5:00
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time only counts
down when solution temperature is equal or greater than set-
point.
>> Detergent solution is recirculated through oscillating jet
system for programmed amount of time. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=) 140.0F
ACIDWASH T= 5:00
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time only counts
down when solution temperature is equal or greater than set-
point.