4-8
122997-341 Operator Manual Washer Operation
>> If pH is too low while injecting, acid neutralizer is injected
to compensate. Display shows:
pH TOO LOW
INJECTING ACIDNEUT.
>> Water-Acid neutralizer solution is mixed. Display shows:
pH TOO LOW
MIXING WATER
>> If pH is too high while injecting, alkaline neutralizer is
injected to compensate. Display shows:
pH TOO HIGH
INJECTING ALK.NEUT.
>> Water-Alkaline neutralizer solution is mixed. Display shows:
pH TOO HIGH
MIXING WATER
>> Water contained in traveling spray header is sent to sump. If
temperature is guaranteed, actual sump temperature
alternates with guaranteed set temperature. Displays shows:
CYCLE 10 (G=) 128.5F
HEADER DRAIN T= 1:00
>> Sump water is pumped to drain. If temperature is
guaranteed, actual sump temperature alternates with
guaranteed set temperature. Display shows:
CYCLE 10 (G=)128.5F
ALK.DRAIN T= 1:00
>> Water still present in sump is gravity-drained (option). If
temperature is guaranteed, actual sump temperature
alternates with guaranteed set temperature. Display shows:
CYCLE 10 (G=)128.5F
GRAVITYDRAIN T= 1:00
>> If water saver is enabled, sump water is sent back to
alkaline tank. If temperature is guaranteed, actual sump
temperature alternates with guaranteed set temperature.
Display shows:
CYCLE 10 (G=)138.7F
ALK.SAVER T= 2:00
>> If sump flush option is available, water still present in
sump is gravity drained. If temperature is guaranteed, actual
sump temperature alternates with guaranteed set
temperature. Display shows:
CYCLE 10 (G=) 128.5F
FLUSH DRAIN T= 00:45
>> If sump flush option is available, water is injected in sump
to remove debris. If temperature is guaranteed, actual sump
temperature alternates with guaranteed set temperature.
Display shows:
CYCLE 10 (G=) 128.5F
FLUSHFILL T= 2:00