EasyManuals Logo
Home>Tektronix>Test Equipment>2230

Tektronix 2230 Service Manual

Tektronix 2230
483 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #180 background imageLoading...
Page #180 background image
Adjustment Procedure— 2230 Service
EXTERNAL Z-AXIS, PROBE ADJUST,
EXTERNAL CLOCK, AND X-Y PLOTTER
Equipment Required (see Table 4-1):
Leveled Sine-Wave Generator (Item 2)
Pulse Generator (Item 5)
Two 50 Q BNC Cables (Item 8)
BNC T-Connector (Item 11)
50 ft BNC Termination (Item 12)
BNC Male-to-Tip Plug (Item 17)
10X Probe (Proveded with Instrument)
INITIAL CONTROL SETTINGS PROCEDURE STEPS
Vertical
1. Check External Z-Axis Operation
Channel 1 POSITION Midrange
a. Connect the leveled sine-wave generator output via
VERTICAL MODE
CH 1
a 50 Q cable and a T-connector to the CH 1 OR X input
X-Y
Off (button out)
connector. Then connect a 50 SI cable and a 50 SI termina-
BW LIMIT
Off (button out)
tion from the T-connector to the EXT Z AXIS INPUT con-
CH 1 VOLTS/DIV
1 V
nector on the rear panel.
CH 1 VOLTS/DIV Variable
CAL detent
Channel 1 AC-GND-DC DC
b. Set the generator to produce a 5 V, 50 kHz signal.
Horizontal
POSITION
Midrange
c. CHECK—For noticeable intensity modulation. The
HORIZONTAL MODE A
positive part ol the sine wave should be of lower intensity
A SEC/DIV
20 ms
than the negative part.
SEC/DIV Variable
CAL detent
XI0 Magnifier
Off (knob in)
d. Disconnect the test equipment from the instrument.
A TRIGGER
VAR HOLDOFF
NORM
Mode
P-P AUTO
2. Check Probe Adjust Operation
SLOPE OUT
LEVEL
Midrange
a. Set:
HF REJECT OFF
CH 1 VOLTS/DIV 10 mV
A&B INT
VERT MODE
A SOURCE
INT
A SEC/DIV 0.5 ms
Storage
b. Connect the 10X Probe to the CH 1 OR X input
STORE/NON STORE
NON STORE (button out)
connector and insert the probe tip into the PROBE
SAVE/CONTINUE
CONTINUE (button out)
ADJUST jack on the instrument front panel. If necessary,
PRETRIG/POST TRIG
POST TRIG (button out)
adjust the probe compensation for a flat-topped square-
ROLL/SCAN
SCAN (button out) wave display.
1K/4K
4K (button out)
POSITION CURS/
POSITION CURS
SELECT WAVEFORM
(button in) c. CHECK—Display amplitude is 4.75 to 5.25 divisions.
WAVEFORM
WAVEFORM
REFERENCE/
REFERENCE
MENU SELECT
(button in)
d. Disconnect the probe from the instrument.
5-32

Table of Contents

Other manuals for Tektronix 2230

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 2230 and is the answer not in the manual?

Tektronix 2230 Specifications

General IconGeneral
BrandTektronix
Model2230
CategoryTest Equipment
LanguageEnglish

Related product manuals