Commands Listed in Alphabetical Order
Syntax
CALibrate?
Examples
CALIBRATE? might return :CALIBRATE:INTER NAL:STATUS
PASS;CALIBRATE:FACTORY:STA TUS PASS0;:CALIBRATE:RESULTS:SP C
PASS;:CALIBRATE:PROBESTATE :CH1 1;CH2 1;CH3 1;CH4 1
, indicating
the calibration status.
CALibrate:CALProbe:CH<x>? (Query Only)
This query-only command instructs the instrument to perform a probe calibration
for the selected channel and returns the calibration status. The Channel <x>
range is 1 through 4. This command is equivalent to selecting Probe Cal from
the Vertical menu. You must warm up the instrument for at least 20 minutes
before running this command.
NOTE. A probe query takes approximately 30 seconds to complete. No other
commands will be executed until calibration is complete.
Group
Calibration
Syntax
CALibrate:CALProbe:CH<x>?
Related Commands
CALibrate:PRObestate:CH<x>?, CH<x>:PROBECal?
Returns
H-1
Indicating the probe calibration failed for the selected channel.
H0
Indicating the probe calibration passed for the selected channel.
H1
Indicating the probe calibration is initialized for the selected channel.
Examples
CALibrate:CALProbe:CH2? might return :CALIBRATE:CALPROBE:CH2 0,
indicating the probe calibration passed for Channel 2.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-103

Table of Contents

Other manuals for Tektronix DPO7000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO7000 Series and is the answer not in the manual?

Tektronix DPO7000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO7000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals