EasyManuals Logo

Tektronix DPO7000 Series Programmer's Manual

Tektronix DPO7000  Series
733 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #251 background imageLoading...
Page #251 background image
Commands Listed in Alphabetical Order
0.00000000000
;:HORIZONTAL:MAIN:INTERPRATI O
1.0000;POSITION 50.0000;UNITS:STRIN G "s";:HORIZONTAL:ROLL
AUTO;MODE AUTO;MODE:AUTO:LIMITRECORDLEN
200000;:HORIZONTAL:MODE:RE CORDLENGTH 1000;SAMPLERAT E
40.0000E+9;SCAL
.
HORizontal:ACQDURATION? (Query Only)
This query
returns the timebase duration.
Group
Horizontal
Syntax
HORizontal:ACQDURATION?
Returns
<NR3>
Examples
HORIZONTAL:ACQDURATION? might return :HORIZONTAL:ACQD
URATION
5.0E-9
, indicating the acquisition duration is 5.0 us.
HORizontal:ACQLENGTH? (Query Only)
This query returns the rec ord length.
Group
Horizontal
Syntax
HORizontal:ACQLENGTH?
Returns
<NR1>
Examples
HORIZONTAL:ACQLENGTH? might return :HORIZONTAL:ACQLENGTH 50000,
indicating the acquisition record length is 50000 points.
HORizontal:DIVisions? (Query Only)
This query-only command returns the number of graticule divisions over w hich
the waveform is displayed.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-239

Table of Contents

Other manuals for Tektronix DPO7000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO7000 Series and is the answer not in the manual?

Tektronix DPO7000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO7000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals