Commands Listed in Alphabetical Order
Examples
TRIGGER:B:TYP
E EDGE
sets the B trigger type to edge.
TRIGGER:B:TYPE? might return :TRIGGER:B:TYPE EDGE, indicating that the
Btriggertype
is set to EDGE.
TRIGger:ENHanced
This command sets or queries the state of trigger position enhancement. When
on, the instrument improves the trigger positioning to more closely match the
acquired data. This is equivalent to selecting Enhanced Triggering from the
Trigger Mode menu.
Group
Trigger
Syntax
TRIGger:ENHanced {<NR1>|OFF|ON}
TRIGger:ENHanced?
Arguments
<NR1> = 0 disables trigger position enhancement, any other value enables trigger
position enhancement.
OFF disables trigger position enhancement.
ON enables trigger position enhancement.
Examples
TRIGGER:ENHANCED ON enables trigger position enhancement.
TRIGGER:ENHANCED? might return :TRIGGER:ENHAN CED 0, indicating that
trigger position enhancement is disabled.
TRIGger:LVLSrcpreference
This command sets or queries the dependent source/level trigger feature.
SRCIndependent is the d efault value.
Group
Trigger
Syntax
TRIGger:LVLSrcpref
erence { SRCDependent|SR CIndependent}
TRIGger:LVLSrcpreference?
Related Commands
TRIGger:{A|B}:LEVel:CH<x>,
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-617