Commands Listed in Alphabetical Order
Arguments
ECL specifies th
epreset
ECL high level.
TTL specifies the preset TTL high level.
<NR3> is the setup and hold data level, in V.
Examples
TRIGGER:A:LOGIC:SETHOLD:D ATA:THRESHOLD TTL specifies the preset TTL
high level as the current data voltage level for the setup and h old trigger.
TRIGGER:A:LOGIC:SETHOLD:D ATA:THRESHOLD? might return
:TRIGGER:A:LOGIC:SETHOLD: DATA:THRESHOLD 1.2000E+00, indicating
that 1.2 V is the current data voltage level for the setup and hold trigger.
TRIGger:{A|B}:LOGIc:SETHold:DATa:THReshold:CH<x>
This command sets or queries the data voltage threshold for setup and hold trigger.
This command is equivalent to selecting A or B Event Trigger Setup from the
Trig menu and then setting the desired Data Level. The value of x can range
from 1 through 4.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGIc:SETHo ld:DATa:THReshold:CH<x>
{ECL|TTL|<NR3>}
TRIGger:{A|B}:LOGIc:SETHo ld:DATa:THReshold:CH<x>?
Related Commands
TRIGger:{A|B}:LOGIc:SETHold:DATa:THReshold:CH<x>
Arguments
ECL specifies the preset ECL high level.
TTL specifies the preset TTL high level.
<NR3> is the setup and hold data level, in V.
Examples
TRIGGER:A:LOGIC:SETHOLD:D ATA:THRESHOLD:CH2 TTL specifies the preset
TTL high level as the current data voltage level for setup and hold trigger f or
Channel 2.
TRIGGER:A:LOGIC:SETHOLD:D ATA:THRESHOLD:CH2? might return
:TRIGGER:A:LOGIC:SETHOLD: DATA:THRESHOLD:CH2 1.2000 E+00,
indicating that 1.2 V is the current data voltage level for setup and hold trigger
for Channel 2.
2-514 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual