Commands Listed in Alphabetical Order
MASK:AUTOSet
:MODe
This command controls whether a mask autoset will be automatically done after
a standard mask is selected. The autoset will never happen if the standard mask
is selected f
rom GPIB, since a “SET?” sent back to the oscilloscope should not
perform a mask a utoset.
The default value is: manual.
This command, like all the MASK:AUTOSET commands, affects only an autoset
on a standard mask, not the general instrument autoset function.
Group
Mask
Syntax
MASK:AUTOSet:MODe {MANual|AUTO}
MASK:AUTOSet:MODe?
Arguments
MANual mask autoset is set manually.
AUTO mask autoset is set automatically.
Examples
MASK:AUTOSET:MODE AUTO sets the mask autoset manually.
MASK:AUTOSET:MODE? might return :MASK:AUTOSET:MODE MANUAL,
i
ndicating that the mask autoset is set manually.
MASK:AUTOSet:OFFSETAdj
This command sets mask autoset not to enforce the rule that, for pulse sta ndards,
require 0 V to be in a certain place in the mask. Instead, mask autoset will attempt
to measure the DC offset in the signal and use o scilloscope offset or position
controls to compensate for the DC offset in the signal.
This command, like all the MASK:AUTOSET commands, affects only an autoset
on a standard mask, not the general instrument autoset function.
Group
Mask
Syntax
MASK:AUTOSet:OFFSETAdj {ON|OFF|<NR1>}
MASK:AUTOSet:OFFSETAdj?
Arguments
ON enables the autoset:offsetadj function.
OFF disables the autoset:offsetadj function.
2-290 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual