TP04100A Interface
&
Applications
Rev
A:
0611
5/98
WARNING
6:
The parts contained inside the thermal wand operate at extremely hot and cold temperatures
unsafe to touch. Do not perform any maintenance inside the thermal wand assembly until the system is turned
off and disconnected. Wait until thermal wand parts have reached a safe and stable temperature near ambient.
WARNING
7:
When connecting thermocouple sensors to the Device Under Test (DUT), electrically isolate
the sensors to protect operators from contact with any DUT site hazardous voltages.
CAUTION
1:
Some of the packing materials in the TP04100A shipment may be a source for Electrostatic
Discharge (ESD) potential. Do not unpack in the vicinity of ESD sensitive components.
CAUTION
2:
Observe the precautions given on the equipment and within this manual to prevent damage to
the equipment.
CAUTION 3:
Before connecting the TP04100A to its electrical source, check that the
-
(ac) voltage and
frequency to be supplied to the TP04100A are correct for those listed on its data plate (located on the rear
panel of the controller cabinet).
CAUTION
4:
Always hold the thermal wand firmly while positioning or re-positioning the stand
arm
CAUTION
5:
Use proper handling and packaging procedures for ESD sensitive circuit boards. Assume that
all circuit boards are the ESD sensitive type.
CAUTION
6:
Unauthorized personnel should not remove from the equipment those panels that are provided
for protection and/or cooling and/or require a tool to remove.
CAUTION
7:
Use suitable Clean Dry Air (CDA) compressed air supply for the TP04100A system a) to
prevent premature fouling of the filterlregulator assembly provided with the system, b) to prevent ice forming
within the cooling module and possibly reducing or obstructing output air flow. Improper air supply quality
can lead to possible damage to system internal operating components.
CAUTION 8:
Properly install, use and maintain the filterlregulator assembly provided with the TP04100A
system. Doing so prevents moisture and/or compressor oils from being introduced into operating components
of the system. If left unchecked, this can cause damage to the system not covered under warranty.
CAUTION 9:
The two thermal runaway Safety sensors (SF1 and SF2), located on the rear panel, must remain
plugged in for normal operation. If SF2 is not plugged in when the system is powered On, then an internal fuse
must be replaced by qualified service personnel before normal operation can resume. Only qualified service
personnel should disconnect SF1 and/or SF2.