Remote Interfaces
TP04100A Interface
&
Applications
Rev A: 0611 5/98
LLIM
Set the lower air temperature limit. Default value: -25
"C
LLIM nnn
--
where
nnn
is -99 to
25
degrees
C.
LLIM?
Read the lower air temperature limit
NEXT Step to the next setpoint during temperature cycling
NOTE:
Stepping will occur whether or not the device is at temperature. NEXT will cause an error if the
system is not in cycling mode.
RMPH
Set the heating ramp rate for the mode currently in use. Default value: Send 0 to set "MAX."
RAMH is a
3
character entry
-
-
where the rate is
1
to 999 in whole increment steps.
RMPH? Read the heating ramp rate
RMPC Set the cooling ramp rate for the mode currently in use. Default value: Send
0
to set "MAX."
RAMH
is a
3
character entry
-
-
where the rate is
1
to 999 in whole increment steps.
RMPC?
Read the cooling ramp rate
RSTP Reset device-specific functions (program mode). The ThermoStream is forced to the
Program Operation Screen. Any device-specific errors are reset
SETD?
Read the dynamic temperature setpoint (changes during a ramp)
SETN Select a setpoint to be the current setpoint
SETN n
--
where n is 0 to 9 if in "Program" mode,
0
-
2 if in "Manual" mode. In "Manual"
mode, setpoint #O is HOT,
#1
is AMBIENT, #2 is COLD.
SEW? Read the current setpoint
#
SETP
Set the current setpoint's temperature. Default values: Factory Settings
SETP nnn.n
-
-
where nnn.n is the temperature (negative sign required for negative
temperatures)
SETP? Read the current temperature setpoint
SFIL Load
a
setup file
SFIL n
-
where n is a #(I-9)
SOAK
Set the soak time for the setpoint currently in use. Default values: 120 (Program); 10
(Manual)
SOAK nnnn
--
where nnnn is 0
-
9999 seconds
SOAK? Read the soak time