EN User manual THERA-Trainer bemo
19
Article no. A007-661_Rev 2.0 - Last edited: 2020-06-02
Plug potential equalization cable into socket 1.
Make sure that the other end of the equipotential bonding cable is connected to the socket of the room.
Make sure that the potential equalisation cable does not get into foot rests or upper body exerciser
accessories during training.
10.2.8. Patient emergency stop / Dummy plug for patient emergency stop
Damage due to incorrect installation!
Make sure that the circular connector is plugged in correctly when connecting the unit.
Note position of plug.
Risk of injury due to damaged safety equipment!
Check that patient emergency stop is functioning correctly before each training session.
Have malfunctions repaired by trained suppliers.
The patient emergency stop is an external button for the patient to immediately terminate a training
session. Ensure that the patient emergency stop can be reached throughout the entire training session.
Training is only possible if a patient emergency stop or dummy plug for the patient emergency stop is
connected. Before training, check whether the patient emergency stop or dummy plug for the patient
emergency stop is connected and working. Otherwise the trainer will display error E556 / E557.
Connect circular connector to socket 1.
Note position of plug.
‒ Note plug guide 2.
‒ Arrow/marking on circular connector points upwards.
Ensure that the cable of the patient emergency stop does not get into footrests or upper body exerciser
accessories during training.