19.3. Cleaning and disinfection
Risk of infection due to contaminated accessories/options!
Wear gloves during cleaning and disinfection.
Clean and disinfect the trainer before each change of patient.
Risk of infection due to contaminated THERA-Trainer!
Wear gloves during cleaning and disinfection.
Follow general precautions to prevent patients/specialists from being exposed to blood-borne pathogens.
Wear personal protective equipment (gloves, masks, lab coats) when contact to bodily fluids such as saliva,
blood, semen or vaginal secretion is likely to occur.
No direct patient contact by health care professionals who have weeping wounds or skin inflammations.
When having direct patient contact as a health care professional who has weeping wounds or skin
inflammations, wear appropriate protective clothing (gloves, scrub caps, etc.).
Danger to life due to electric shock!
Prior to any cleaning or disinfection, disconnect the power plug from the socket.
Make sure that no cleaning agent or disinfectant gets into the THERA-Trainer.
If any cleaning agent or disinfectant gets into the THERA-Trainer, contact a trained supplier immediately.
Material damage due to incorrect cleaning or disinfection.
Only use disinfectant wipes.
Never use cleaning agents that are aggressive, abrasive or caustic, or contain solvents.
Use only (alcohol-based) disinfectants that comply with the hygiene and disinfection schedule of the facility:
– When selecting, check the relationship between concentration and time. Only use fast-acting
disinfectants.
– Use liquid disinfectants or moist, germicidal disposable wipes.
– Note effectiveness with increasing ranking of alcohol-based disinfectants: methanol < ethanol <
isopropanol < n-propanol.
Only use mild and environmentally friendly cleaning agents and disinfectants.
Follow manufacturer safety instructions for cleaning agents and disinfectants.
Cleaning and disinfection of the THERA-Trainer:
Local conditions and the assessment of the risk of infection form the basis for determining the scope of the
necessary cleaning and disinfection measures.
Clean/disinfect the THERA-Trainer regularly in accordance with the hygiene and disinfection schedule of the
facility.
Use only a damp and soft cloth for cleaning and disinfection.
Use only (alcohol-based) disinfectants that comply with the hygiene and disinfection schedule of the facility.