EN User manual THERA-Trainer bemo
47
Article no. A007-661_Rev 2.0 - Last edited: 2020-06-02
Cleaning and disinfection:
Disconnect power plug.
Check electronic components for damage.
– If any damage is found, do not clean but contact trained supplier.
Make sure that the room is adequately ventilated during cleaning and disinfection.
Use disinfectant wipes on all surfaces with frequent hand/skin contact after each use.
– Clean/disinfect foot rests, arm rests and all options required for therapy.
– Wipe foreign liquids (blood, urine etc.) with disposable germicidal wet wipes, or disinfect the device
using an alcohol-based disinfectant wipe.
– Use disinfectant wipes on all contact surfaces (control and display unit, therapy grips/arm rests, foot
rests, etc.) before and after every use, especially when the device is used by patients with broken skin
or incontinence problems.
– Alternatively it is also possible to protect the contact surfaces with sterile disposable bags or foils,
which are to be changed after every use.
Perform targeted disinfection activities in the event of identifiable contamination, an outbreak situation
or the occurrence of special pathogens.
– Use a sufficient amount of disinfectant and rub the surfaces under light pressure.
Clean other surfaces as required using water with additives that boost cleaning power.
Carefully clean and disinfect foils and stickers to avoid damaging them.
Allow cleaned/disinfected parts to dry completely.
Only use the THERA-Trainer again after the contact time and drying time defined by disinfectant
manufacturer has passed.
Follow the instructions of the disinfectant manufacturer, as well as applicable legal provisions.
Cleaning and disinfection of the velcro straps:
Replace velcro straps if they have been contaminated with blood or secretions.
Cleaning by machine washing with household detergent is possible as it is a non-critical medical device.
For disinfection, place velcro straps into a disinfectant bath.
Allow cleaned/disinfected velcro straps to dry completely.
Only use the velcro straps again after application time and drying time defined by disinfectant manufacturer
has elapsed.